Microphone Check
Use the Microphone Channel, Microphone Quality, and Beamforming Quality menus to check the device’s microphones and beamforming results. Use them immediately after installation or when measurements seem questionable.
The three menus serve these purposes:
- Microphone Channel — View the waveform and spectrum of one microphone channel. You can also turn the 40 kHz reference signal on and off here.
- Microphone Quality — Measure all 112 channels at once to check sensitivity differences and noise across channels.
- Beamforming Quality — Collect beamforming results over a set period and check the difference between the maximum and minimum values.
※ Check results are not saved on the device. Both administrator and regular user accounts can perform these checks.
Microphone Channel
The Microphone Channel menu (titled “Microphone” on the screen) shows the signal from one selected microphone channel.

Mic ID
Enter the microphone number you want to check in the “Mic ID [0, 111]” field. There are 112 microphones, numbered 0–111. The factory default is 88.
- Entering a value within the range changes the selection immediately and displays a result such as “Mic ID=88”.
- Entering a value outside the range displays “id must be in the [0, 111] range” and does not send it to the device.
- The selected number is saved on the device and persists after a reboot.
※ Mic ID only determines which channel is shown on this screen. It does not affect beamforming results.
Waveform
Shows the selected channel’s signal over time. The horizontal axis shows seconds, and the vertical axis shows signal amplitude. The most recent one-second interval is displayed. The vertical axis adjusts automatically to the signal amplitude.
Spectrum
Shows the selected channel’s level (dB) by frequency. The browser calculates the values by analyzing approximately 0.3 seconds of signal at a time, then smooths them with exponential averaging.
Spectrum options apply only to the browser screen. They are not saved on the device and return to their defaults when you reopen the screen.
- MIN Freq, MAX Freq — Integers from 0 to 100000 Hz. Defaults: 0 and 100000. MIN must be less than MAX.
- MIN Y-Axis, MAX Y-Axis — Integers. Defaults: -50 and 50. MIN must be less than MAX.
- Exponential averaging — Positive integer. Default: 10. A value of 1 shows only the most recent values without averaging; larger values make the curve smoother.

Reference
Click “Add Reference” to fix the current spectrum curve as a red line for comparison with subsequent curves. Click “Delete Reference” to remove it.
- While it is fixed, “Reference Set” appears in red.
- If there is no spectrum to display yet, “No spectrum data to capture” appears.
※ The reference is retained only on this screen. It disappears if you refresh the page or move to another menu.
40 kHz Reference Signal
The device can generate a 40 kHz signal for testing. Turn it on and off under “40KHz Signal Generator Control” in the Microphone Channel menu. Use it to check whether the microphone detects 40 kHz sound.
“Current Status” shows one of the following:
- ON — The signal is on. The button says “Turn Off”.
- OFF — The signal is off. The button says “Turn On”.
- UNKNOWN — The status has not yet been read. The button cannot be pressed.
The status is read once when you open the screen. If it is changed elsewhere, refresh the page to see the new status. If UNKNOWN persists, check communication with the device and refresh the page.

When the signal is on, the device measures this 40 kHz signal instead of actual sound. The measurement band must include 40 kHz for the microphone to detect the signal.
Turn the signal off as soon as the check is complete. If left on, this signal will be measured instead of actual sound, affecting beamforming results and the level trigger.
Microphone Quality
The Microphone Quality menu measures the signal amplitude (RMS) of all 112 channels at once. Use it to find channels with different sensitivity or high noise.

1. Set the Measurement Conditions
- Measurement Duration (seconds) — 0.1–0.5; default: 0.5. An out-of-range value displays “Measurement duration: 0.1–0.5”.
- 40kHz Reference Signal — Off by default
The measurement method depends on the switch setting.
- On — Turns on the reference signal and measures in the 38 kHz–42 kHz band. Use this to check sensitivity differences across channels.
- Off — Turns off the reference signal and measures using the current measurement band. Use this to check the noise floor of each channel.
※ When measuring with the switch on, the device changes the measurement band to 38 kHz–42 kHz, waits one second, then measures. Afterward, it restores the original band and reference signal state.
※ Changes to the measurement band made elsewhere while a measurement is in progress wait until the measurement is complete.
2. Run
Click “Run”. The button displays “Measuring…” while the measurement is in progress.
You cannot run it while AI is operating. The screen displays “Microphone calibration cannot run while AI is operating.” and disables the button. Turn off AI or run it outside its operating hours.
※ AI status is checked once when you open the screen. If AI starts after you open it, refresh the page.
3. Check the Results
When the measurement is complete, “Complete: 112 channels” appears, followed by a graph and table.
- Graph — The horizontal axis shows channels, and the vertical axis shows RMS. Three lines show min, max, and avg.
- Table — Contains Index, min, max, and avg columns. Values are displayed to two decimal places.
- If the measurement fails, “Failed: …” or “Error: …” appears.
A channel that receives no signal is marked “No signal” in the table and appears as a gap in the graph. If no channel receives a signal, “No signal detected on any channel.” appears above the graph.

How to Read the Results
- With the reference signal on, an avg value noticeably lower or higher than those of nearby channels may indicate a sensitivity issue in that channel.
- With the reference signal off, a channel with a higher avg value than nearby channels has more noise.
- A “No signal” channel may indicate a microphone or connection problem.
- Save the results from immediately after installation as a baseline for comparison with later checks.
Beamforming Quality
The Beamforming Quality menu collects beamforming results over a specified period and calculates statistics. Use it to check the difference between the maximum and minimum beamforming results numerically.

1. Choose the Measurement Duration
Under “Measurement Duration”, choose 1–5 seconds. The default is 3 seconds.
2. Start
Click “Start”. During collection, the button displays “Collecting…” and the measurement duration selector is locked.
3. Check the Results
When collection is complete, a graph and three cards appear. The graph title shows the number of frames collected.
- Graph — Plots MIN (blue), MAX (red), and AVG (green) for each frame. The values are calculated from the 1,144 cells of the beamforming result.
- AVG(MAX) — Average of the per-frame maximum values
- AVG(MIN) — Average of the per-frame minimum values
- DIFF — AVG(MAX) minus AVG(MIN)
Values are in dB and displayed to three decimal places.
※ Compare these values only when the measurement band, ambient sound, and device position are the same. Values can differ even for the same device if conditions change.
※ This check can run regardless of whether AI is operating.